BD CSampler |
: | Detailed Specifications |
| Power Requirements | : | No additional power necessary |
| Software Requirements | : | BD CFlow Sampler Software |
| Compatible Plate Types | : | Standard 96-well (flat, round, and v bottom) plates Deep-well 96-well plates 48-well plates 12 x 75 mm tubes can also be accommodated, using 24-tube rack supplied. Plates manufactured to the guidelines published by the American National Standards Institute, submitted by the Society for Bimolecular Screening, should be compatible with BD CSampler. |
| Space Requirements | : | Minimum bench depth 28" (71 cm) Width of BD Accuri C6 with BD CSampler installed is 19.5" (49.5 cm) |
| Processing Time | : | <90 minutes for 96-well plate, utilizing 30 second acquisition time per well |
| Weight | : | 7 lbs (3.2 kg) |
| Minimum Sample Volume | : | Does not exceed 50 µL for 96-well plates |
| Wash Cycle | : | Up to 3 wash cycles per well |
| Agitate Cycle | : | Up to 3 agitate cycles per well Single agitate cycle will maintain fixed lymphocytes from human peripheral blood in suspension (+/-10% of expected value) while processing a 96-well, round-well plate, 30 seconds data acquisition per well. |