Always follow the startup and shut down procedures outlined in the BD Accuri C6 flow cytometer Instrument Manual in order to keep the SIP and flow cell clean, and the fluidics system free of air bubbles. Decontamination fluid is run through the BD Accuri C6 fluidics system at each shut down. This helps to keep the SIP and flow cell free of debris and protein build-up but may not prevent build-up from occurring.
We provide a variety of products for day-to-day operation such as Cleaning, Decontamination and Bacteriostatic Concentrates. We have packaged many of these items into convenient kits.
Run our Validation Beads each day the BD Accuri C6 is used. Laboratories that intermittently use their BD Accuri C6 to perform cell counts should validate performance prior to each set of counting experiments. Absolute count measurements can only be performed in BD CFlow Plus or BD CFlow Sampler software.